Home>BS Standards>BS EN IEC 62961:2018 pdf download

BS EN IEC 62961:2018 pdf download

BS EN IEC 62961:2018 pdf download.Insulating liquids – Test methods for the determination of interfacial tension of insulating liquids – Determination with the ring method.
7.3 Determination of the surface tension of water used for the test Start the measurement according to the manufacturer instfuctions.
Introduce water into a clean vessel to a level of 20 mm to 30 mm. Slowly raise the platform and the measurement vessel until the ring is immersed In the water phase. The ring shall be at least 3 mm below the water surface and centred in the vessel. Ensure that the ring is properly wetted and not repelled from the water surface.
Depending on the test temperature, a value of 70 mNIm to 73 mNlm shall be obtained. If lower values are found, repeat cleaning of the sample vessel and the ring. Check the source of the sample water in respect to surface active contaminants. In exceptional cases, search for a fresh sample of water from another source (see also 6,3).
If the water quality (according to 8.3) and the cleanliness of vessels (according to 6.1) and the ring (according to 6.2) are validated in the testing laboratory by an established and reproducible procedure, it Is not necessary to carry out the determination of the surface tension of water (according to 7.3) prior to each sample, but once in a daily successive series of tests.
7.4 Determination of interfacial tension between water and insulating liquid
Introduce water into a clean vessel to a level of 20 mm to 30 mm. Slowly raise the platform and the measurement vessel until the ring is immersed in the aqueous phase. The ring shall be at least 3 mm below the water surface and centred in the vessel.
• Then overlay the water with a 15 mm to 20 mm layer of insulating liquid using a pipette or a clean narrow beaker with spout (if plastic single use syringes are used it shall be verified that the material does not interact with the liquid). No bubbles shall be present. Complete this process within 30 s.
• Begin the time counting.
• Measurement shall be completed within 180 s t 30 s.
• The measurement value shall be taken at the time closest to 180 s (this time shall not differ from 180 s by more than 30 s).
• Ascertain whether the obtained value is to be corrected or if the coi’rection factor has already been implemented in the software.
NOTE I Some instruments can automaiie the procedures above.
NOTE 2 A secies of several conseculive measurements on the same sample interface (diffefent surface age) witrnn the given time frame can be made
8 Test report
The test report shall include at least the following information:
• a reference to this document and to the method used:
• identification of test specimen:
• the date and place where the determination has been carried out:
• the type of correction formula, if other than that of Zuidema and Waters 151. as well as further significant deviations from the described procedure:
• the value of the measured interfacial tension at approximately 180 s according to 7.4:
• ring speed in mm/s (Optional).
Some laboratories have evaluated other methods of testing the IFT of insulating liquids. The drop volume method is a particularly reasonable alternative to the force (ring) method as it overcomes some of those weaknesses. In the tests carried out within the round robin tests, fast and reproducible results at a well-defined interface age could be obtained. It allows multiple tests to be carried out with many fewer samples and without the need for refilling. The parts in contact with insulating liquid do not need flame cleaning. Syringes and needles are stable and robust.
It shall, however, be considered that the result of the drop volume method is more sensitive to the exact determination of the density of the liquid than that of the ring method.
A.2 Principle of the method
A.2.1 Basics
The interfacial tension between insulating liquid and water can be determined with a drop volume tensiometer. The method uses drops (water or insulating liquid, depending on the instrument configuration) produced by means of a vertically positioned capillary using a suitable, high-precision dosing device.
The size of each droplet increases with specific volume flow until they break off from the dosing capillary. The detached droplets are detected by a suitable sensor. When the volume flow and the number of droplets are known, the break-off volume for each droplet can be calculated. This measurement of break-off volume enables calculation of the surface tension from the balance of forces at the moment of break-off. In the case of interfacial tension, the needle Is dipped into the test liquid — either Insulating liquid or water — depending on the instrument configuration. The droplet will detach at the moment when the buoyancy balances the wetting force at the capillary tip.

Related Standards